REV 2 IRM-5000P USER’S MANUAL
17
f. The following screen will be displayed:
1. YES
Press the
[1]
key (YES) to print the resistance readings on the IRM-5000P’s built-
in thermal printer. If this option is selected, the resistance values will be printed 
at every second, starting at 10 seconds. Continue to step g. 
2. NO
Press the
[2]
key (NO) if you do not want to print the resistance readings.
Continue to step g.
g. The following screen will be displayed if the IRM-5000P needs to perform self-
calibration:
The remaining calibration time is displayed in seconds. When the calibration is 
completed, the following screen will be displayed: 
Press the
[ENTER]
key to start the test.
h. The following screen will be displayed while the test is being performed:
NOTE
The resistance value symbols are: 
T = Tera-ohm 
G = Giga-ohm 
M = Mega-ohm 
K = Kilo-ohm 
Bat:
Time: 09:51
Resistance: 983.8 M 
Current: 47 uA 
TEST VOLTAGE: 4999V 
 “ENTER” TO START!
R 5000V 10Min 
 
TERMINAL VTG: 1 V 
 CALIBRATING 
 PLEASE WAIT 25 
 
TERMINAL VTG: 1 V 
PRINT DURING TEST?
1. YES 2. NO 
 
TERMINAL VTG: 1 V