Vanguard IRM-5000P User Manual
Page 34
 
IRM-5000P USER’S MANUAL
REV 2
30
g. The following screen will be displayed while the test is being performed:
The LCD screen will be dynamically updated displaying the remaining time along with 
the resistance, current, and test voltage values. When the test is completed, the results 
for the first time interval and voltage step will be displayed as shown: 
Press any key to continue. The results for the second time interval and voltage step will 
be displayed: 
Press any key to continue. The results for the third time interval and voltage step will be 
displayed: 
Press any key to continue. The results for the fourth time interval and voltage step will 
be displayed: 
Press any key to continue. The results for the final time interval and voltage step will be 
displayed: 
Press any key to continue.
h. The following screen will be displayed:
 9.863G@ 05:00, 2500V
2500V 0.00uF 05:00 
 
TERMINAL VTG: 1 V 
 9.847G@ 04:00, 2000V
2500V 0.00uF 05:00 
 
TERMINAL VTG: 1 V 
9.851G@ 03:00, 1503V
2500V 0.00uF 05:00 
 
TERMINAL VTG: 1 V 
 9.973G@ 02:00, 1002V
2500V 0.00uF 05:00 
 
TERMINAL VTG: 1 V 
 9.952G@ 01:00, 499V
2500V 0.00uF 05:00 
 
TERMINAL VTG: 1 V 
Bat:
Time: 04:59
Resistance: 10.002 G 
Current: 50.2 nA 
TEST VOLTAGE: 498 V