Vanguard IRM-5000P User Manual
Page 38

IRM-5000P USER’S MANUAL
REV 2
34
3.1.5. Performing a Dielectric Discharge (DD) Test
The dielectric discharge (DD) test measures the dielectric absorption of a tested material. A
typical DD test is timed for 30 minutes. Follow the steps below to perform a dielectric discharge
test:
a. Start from the “START-UP” menu:
Press the
[1]
key (RUN TEST).
b. The following screen will be displayed:
Press the
[3]
key (DD).
c. The following screen will be displayed:
Press the
[1]
key (30 Min).
d. The following screen will be displayed:
Press the
[1]
key to select 500V or press the
[2]
key to enter a custom test voltage
value.
1. 500V
2. Custom
TERMINAL VTG: 1 V
1. 30 Min (Default)
2. Custom
TERMINAL VTG: 1 V
1. R 2. PI
3. DD 4. SV
TERMINAL VTG: 1 V
1. RUN TEST 05/05/10
2.
UTILITIES 10:15:15
Bat:
TERMINAL VTG: 1 V