How SpO

2

Information is Displayed

The R Series pulse oximeter displays the following information in the SpO

2

area:

SpO

2

value (arterial oxygen saturation percentage)

SpO

2

alarm indicator

Signal strength indicator (the bar rises and falls to indicate the relative change in the
pulsatile signal)

The display optionally includes the normalized plethysmogram below the ECG trace. The
R Series pulse oximeter calculates oxygen saturation and updates the display once per second.

Note:

If you are using the Plus or BLS models, you will see the SpO2 value, the alarm
indicator, and the signal strength indicator while in AED or Manual mode. To see the
normalized plethysmogram, the unit must be in Manual mode.

Monitoring a Patient

To set up the pulse oximeter:

1. Attach the sensor to the patient, and plug the sensor cable into the patient cable (refer to

“Applying a Reusable DCI Sensor” on page 6 or “Applying a Single-Use Sensor” on
page 7.
)

2.

Ensure the patient cable is connected to the unit.

3.

Turn the Mode Selector to the following:

MONITOR for ALS models.

ON for Plus or BLS models. Press the Manual Mode softkey, and then the Confirm
softkey to enter the Manual mode of operation.

When the unit obtains measurements for arterial oxygen saturation, numeric values replace
the dashed lines on the display.



6S2

Options

Param

Code

Marker

Report

Alarms

11:24

ECG

SpO2

81

ECG

PADS

x1

SpO2%

CPR

0:00

IDLE

Depth
Rate

99

Test
Log

Plethysmogram

Signal

Strength

Indicator

SpO

2

Alarm Status

Arterial Oxygen

Saturation Percentage

(or dashed line if no
valid measurement)