Specifications

The following specifications apply to the R Series pulse oximeter only. For information about
the R Series defibrillator and batteries, refer to your defibrillator’s R Series Operator’s Guide.

CHECK SPO2 SITE

Insufficient perfusion at the sensor
site

Reposition or relocate the
sensor, or increase perfusion.

CHECK SPO2 SENSOR

An inappropriate sensor site, poor
application of the sensor, no
patient cable plugged in, or motion
is causing invalid SpO

2

readings.

Reposition or relocate the
sensor, or increase perfusion.

Ensure that the cable is
plugged into the unit.

Cease motion.

Dashes appear in place
of the SpO

2

saturation

percentage.

Excessive ambient light,
inadequate perfusion, high signal
artifact, or a defective or
disconnected sensor or patient
cable

Reposition or relocate the
sensor, reduce the ambient
light, shield the sensor from the
light, or increase perfusion.

Check the cable and sensor.

Message:

SPO2 COMM ERROR

The pulse oximeter subsystem is
not communicating with the
defibrillator.

Turn unit off and then back on
to see if message clears. If
message persists, contact
ZOLL Technical Service.

Message or Symptom

Possible Causes

Recommended Action

General

Saturation range (%SpO

2

)

1% to 100%

Saturation resolution

1%

Saturation accuracy *

Note: R Series defibrillators are

designed to work with the
Masimo LNCS series of
sensors and are dependent on
their specifications. Refer to
Masimo LNCS sensor labeling
for accuracy information.

During no-motion conditions

Adults/pediatrics:

70% to 100%, ±2%

0% to 69%, unspecified

Neonates:

70% to 100%, ±3%

0% to 69%, unspecified

During motion conditions

All patients:

70% to 100%, ±3%

0% to 69%, unspecified

Saturation alarm limits

On/Off displayed on monitor; operator-selectable.

Low limit:

70% to 98%

High limit:

72% to 100%

SpO

2

Wavelength

Nominal Red LED Wavelength:

660 nanometers

Nominal Infrared LED Wavelength: 905 nanometers

Energies (Radiant Power) of light for
LNCS Sensors at 50 mA pulsed

Minimum:

0.13 mW

Maximum:

0.79 mW

Pulse rate range

25 to 240 beats per minute

Pulse rate resolution

1 beat per minute