Spo2 accessories, Messages and troubleshooting, Accessories – ZOLL R Series Monitor Defibrillator Rev D User Manual

Page 17

Advertising
background image

SpO2 Accessories

9650-0901-01 Rev. D

R Series Pulse Oximetry (SpO

2

)

15

appear after the unit is powered-up. See the R Series Configuration Guide for instructions on
changing power-up defaults.

SpO

2

Accessories

The following table describes each of the SpO

2

accessories.

Messages and Troubleshooting

Parameter

Factory Default Setting

Possible Values

Averaging period

8 seconds

4 seconds

8 seconds

16 seconds

Sensitivity level

Normal

Normal

High

SpO

2

low saturation limit

85%

50% to 100% or OFF

SpO

2

high saturation limit

OFF (appears as: ---)

50% to 100% or OFF

Low Heart Rate Alarm Limit

30 beats per minute

20 to 100 beats per minute

High Heart Rate Alarm Limit

150 beats per minute

60 to 280 beats per minute (ECG)

60 to 235 beats per minute (SpO

2

)

Item

Description

REF

LNCS Adtx

Single use sensor for Adult patients > 30 kg

8000-0320

LNCS Pdtx

Single use sensor for Pediatrics and Slender Adults 10 - 50 kg

8000-0321

LNCS Inf-L

Single use sensor for Infants 3 - 20 kg

8000-0322

LNCS Neo-L

Single use sensor for Neonates < 3 kg

8000-0323

LNCS NeoPt-L

Single use sensor for Neonates < 1 kg (Pre-term)

8000-0324

LNCS DCI

Reusable sensor for Adults and Pediatrics > 30 kg

8000-0294

LNCS DCIP

Reusable sensor for Pediatrics 10 - 50 kg

8000-0295

LNC-4

4' Reusable Patient Cable

8000-0298

LNC-10

10’ Reusable Patient Cable

8000-0293

LNC Ext

LNC Extension Cable, DB-9 Termination, 4ft

8000-0325

LNCS-to-LNOP

Adapter Cable, LNCS Sensor to LNOP Patient Cable

8000-0327

LNOP DC-12

LNOP Adult Reusable Direct Connect 12’ Cable

8000-0296

Message or Symptom

Possible Causes

Recommended Action

SPO2 AMBIENT LIGHT

Excessive ambient light

Relocate the sensor, reduce
the ambient light, or shield the
sensor from the light.

SPO2 PULSE SEARCH

Sensor cannot detect a pulse.

Normal behavior immediately
after applying the sensor. If this
message persists beyond 10
seconds, reposition or relocate
the sensor, or increase
perfusion.