Daily Test Procedure

Perform the following procedure daily to ensure that the pulse oximeter is functioning properly.

Use a reusable oximeter sensor for this procedure and test the pulse oximeter as follows:

1. Attach a reusable SpO

2

sensor to your finger and connect the patient cable to the SpO

2

connector.

2.

Turn the Mode Selector to the following:

MONITOR for ALS models.

ON for Plus or BLS models. Press the Manual Mode softkey, and then the Confirm
softkey to enter the Manual mode of operation.

The monitor displays the pulse signal indicator and the arterial oxygen saturation
percentage.

3.

Manually measure your pulse and compare it with the displayed pulse rate.

The displayed pulse rate should be comparable to your measured pulse.

4.

Observe the plethysmogram and verify that the wave repeats at the pulse rate.

If the unit is not displaying the waveform, refer to “Displaying the Plethysmogram” on
page 12.

5.

With SpO

2

alarms enabled, raise the low SpO

2

limit to generate an alarm, suspend the alarm

tone, and then reset the low limit. (For instructions, refer to “Setting SpO2 Alarms” on
page 13.)

When the alarm occurs, the unit emits a continuous alarm tone, highlights the SpO

2

value,

and flashes the associated bell symbol.

6.

Lower the high SpO

2

limit to generate an alarm, suspend the alarm tone, and then reset the

high limit. (For instructions, refer to “Setting SpO2 Alarms” on page 13.)

When the alarm occurs, the unit emits a continuous alarm tone, highlights the SpO

2

value,

and flashes the associated bell symbol.

7.

Remove the sensor from your finger without disconnecting the patient cable.

The unit emits two beeps and displays the message: CHECK SPO2 SENSOR.

8.

Replace the sensor on your finger and wait for the SpO

2

value to reappear.

9.

Unplug the patient cable from the unit.

The unit emits two beeps and displays the message: CHECK SPO2 SENSOR.

Default Settings for the SpO

2

Option

The table below lists the factory default settings for SpO

2

monitoring and the range of values

available for each parameter. Unless changed by the user, these default settings will always