Specifications

9650-0901-01 Rev. D

R Series Pulse Oximetry (SpO

2

)

17

Note:

The Pulse Oximetry Option is calibrated for functional saturation.

Note:

Because R Series pulse oximeter measurements are statistically distributed, only about
68% of these measurements can be expected to fall within plus or minus one standard
deviation of the value measured by a CO-oximeter.

* The accuracy of the R Series pulse oximeter can only be verified by comparing its
measurements to those obtained with a CO-oximeter. Pulse oximeter functional testers cannot
be used to verify the accuracy of the oximeter or its sensors.

Pulse rate accuracy

During no-motion conditions

25 to 240 beats per minute, ±3 beats per minute

During motion conditions

25 to 240 beats per minute, ±5 beats per minute

Pulse rate alarm limits

On/Off displayed on monitor; operator-selectable.

Low limit:

20 to 100 beats per minute

High limit:

60 to 235 beats per minute

Biocompatibility

Patient-contacting material meets requirements of
ISO 10993-1, Biological Evaluation of Medical Device,
Part I, for external surfaces, intact surfaces, and
short-term exposure.

Environmental

Temperature

Operating:

0°C to 40°C

Storage and shipping:

–20°C to 60°C

Note: The defibrillator might not perform to

specifications if stored at the upper or lower
temperature extreme and then put to immediate
use.

Electromagnetic immunity

EN60601-2-4; IEC 1000-4-3 to 18 V/m

Software Hazards

Minimized by compliance with EN14971