Setting SpO

2

Alarms

The R Series pulse oximeter provides operator-programmable alarms for arterial oxygen
saturation and pulse rate. These are high-priority alarms, indicated visually by flashing the
associated bell symbol and parameter values and audibly by a continuous tone.

Note:

Once the arterial oxygen saturation reaches the high or low limit, there is a 4 second
delay until the alarm occurs.

For the low and high alarm limit ranges and factory default values, refer to “Specifications” on
page A-1 of your defibrillator’s R Series Operator’s Guide.

Note:

When monitoring heart rate via pulse oximetry rather than ECG electrodes, the
tachycardia alarm limit is automatically lowered to 235 if it was previously set to a
higher value. The unit restores the higher setting when ECG monitoring resumes.

SpO

2

Automated Alarm Limits

When the SpO

2

alarm state is set to AUTO, the unit calculates the low and high limits for

arterial oxygen saturation as follows:

To use AUTO alarm limits, ensure that the unit is making valid SpO

2

measurements from the

patient.

To set SpO

2

alarm limits:

1. Using the procedure “Setting Alarms for Monitored Parameters” in your defibrillator’s

R Series Operator’s Guide, select the parameter SpO

2

.

2.

Follow procedures in the Alarm section of your defibrillator’s R Series Operator’s Guide to
enable or disable processing and/or set alarm limits.

For information on configuring different alarm default values, refer to the R Series
Configuration Guide.

.

SpO

2

Alarm Limit

(AUTO State)

Calculation

Low

95% of the patient’s current saturation measurement

High

105% of the patient’s current saturation measurement (up to 100%)

Inc

>

Dec

<

Enter

Parameter

State

Low

High

SpO2

ENABLE

ENABLE

30

85

150

100

ECG HR

Cancel

81

ECG

II

x1

SpO2%

CPR

0:00

IDLE

Depth
Rate

99