<8. Device Information>

8-7

IM 01C25T02-01E

Table 8.12

Action of each parameters in failure mode related Advanced Diagnostic (option code: /DG1)

ALARM

Display

Cause of Alarm

SENSOR Transducer block

DIAG_ERR

PV.STATUS

SV.STATUS

TV.STATUS

(*1)

Hexadecimal

Display through DD

AL.87

FLG HI

Flange Temperature High Alarm occurs.

0x4000

Flg Temp High Alarm

AL.87

FLG LO

Flange Temperature Low Alarm occurs.

0x2000

Flg Temp Low Alarm

AL.88

INVR.DP

Reference fDP value is invalid.

0x0800

Invalid Ref fDP

AL.88

INVR.SL

Reference fSPl value is invalid.

0x0400

Invalid Ref fSPl

AL.88

INVR.SH

Reference fSPh value is invalid.

0x0200

Invalid Ref fSPh

AL.88

INVR.F

Reference BlkF value is invalid.

0x0100

Invalid Ref BlkF

AL.89

B BLK

B Blocking is detected.

0x0080

B Blocking

UNCERTAIN:Non Specific

AL.89

H BLK

High-pressure-side Blocking is detected.

0x0040

High Side Blocking

UNCERTAIN:Non Specific

AL.89

L BLK

Low-pressure-side Blocking is detected.

0x0020

Low Side Blocking

UNCERTAIN:Non Specific

AL.89

H LRG

Large Fluctuation of High Side.

0x0010

Large Fluctuation of High Side UNCERTAIN:Non Specific

AL.89

L LRG

Large Fluctuation of Low Side.

0x0008

Large Fluctuation of Low Side

UNCERTAIN:Non Specific

AL.89

A BLK

A Blocking is detected.

0x0004

A Blocking

UNCERTAIN:Non Specific

AL.89

DIAG.OV

DPavg is outside range.

0x1000

Outside Diagonosis Range

(*1): These settings depend on bit 15 in DIAG_OPTION.