<Appendix 8. Advanced Diagnostics>

A8-8

IM 01C25T02-01E

A8.2.5 Alarm and Alert Setting

The abnormal results, as the blockage detection

and high/low flange temperature (heat trace

monitoring) are given by an analog alert or the LCD

display of alarm status. The analog alert or the LCD

display of alarm status is set according to the flow

shown in below figure.

FA0806.ai

Discrete Alarm

FF Alert Notify and

(STB.FLG_TEMP_ALM)

Outside Diagnosis Range/

Invalid Ref xx

Masking

PV/SV/TV Status

Reflection Flag

(STB.DIAG_OPTION)

PV/SV/TV Status

Alarm Display

on LCD

Discrete Alarm

FF Alert Notification

(STB.DIAG_H_ALM)

(STB.DIAG_L_ALM)

Storage of Abnormal results

(STB.DIAG_ERR)

Alarm Masking

(STB.DIAG_OPTION)

Device Status

(RB.DEVICE_STATUS_7)

Alarm Enabling

(STB.ALAM_SUM.DISABLED)

Report Setting

(STB.DIAG_PRI)

Figure A8.5 Alarm and Alert Setting

A8.2.5.1 Alert Setting

Link to DIAG_H_ALM and DIAG_L_ALM

DIAG_H_ALM and DIAG_L_ALM enable to indicate

the alarm status separately divided into the high-

and low-pressure-side alert. The factory setting is

not enabled to display them.
Enable to display the alarm status to DIAG_H_ALM

and DIAG_L_ALM according to the following

procedure.

1) Set "Stop" to DIAG_MODE.

2) Uncheck the checkbox of "Diag Alm Disabled”,

which is corresponding to bit 8, in ALARM_

SUM.

Note: Set to “Calculation” after setting the parameters.

Report Setting

DIAG_PRI in SENSOR Transducer block defines

the priority to transfer the analog alert to the host.

The alert is generated according to the priority when

interfering in other alerts.

DIAG_PRI

Bit

Description

0

An alert is not generated.

1

An alert does not be reported to the host.

2

Disabled

3-7 ADVISORY

8-15 CRITICAL

The default value at the factory setting is set to 1.
Set the priority according to the following procedure.

1) Set "Stop" to DIAG_MODE.

2) Enter the value to DIAG_PRI in the number of 3

to 15.

Note: The setting of the highest priority is 15.

Set to “Calculation” after setting the parameters.

A8.2.5.2 Alarm Display on LCD

If the ILBD algorithm detects the abnormality, the

content of the detected result is displayed with

“AL.88” or “AL.89” on the LCD. “AL.88” indicates

that condition is not applicable for the abnormality

detection and “AL.89” indicates the abnormality is

detected.

FA0808.ai

Figure A8.6 Display Example of High Side Blocking

The alarm display on LCD regarding ILBD is

described in Table 8.12 of the section 8.

A8.2.5.3 Alarm Status

When the algorithm of ILBD and Heat trace

monitoring detect the abnormality, the result is

stored in DIAG_ERR. The alarm status based on

the detected abnormality is displayed to DIAG_ERR

in the SENSOR Transducer block.

DIAG_ERR

Bit

Alarm status

D

IA

G

_H

_

A

LM

D

IA

G

_L

_

A

LM

FL

G

_

TE

M

P_

A

LM

0 Not used.
1 Not used.
2 A Blocking

3 Large Fluctuation of Low Side

4 Large Fluctuation of High Side