<Appendix 8. Advanced Diagnostics>

A8-14

IM 01C25T02-01E

A8.2.12 ILBD Parameter Lists

All the ILBD parameters belong to the SENSOR Transducer block.
Note: The Write Mode column contains the modes in which each parameter is write enabled.

O/S:

Write enabled in O/S mode.

MAN: Write enabled in Man mode and O/S mode.

AUTO: Write enabled in Auto mode, Man mode, and O/S mode.

Relative

Index Index

Parameter Name

Factory

Default

Write

Mode

Explanation

66

2066 DIAG_MODE

Stop (0)

AUTO Select the function mode for the ILBD.

Stop (0):

The blockage detection is stopped.

Calculation (1):

The blockage detection is carried out. The alarms

are generated along with the detected result.

Reference (2):

The reference values are obtained and the update

values are overwritten. After setting, this mode

changes automatically to Calculation (1)

67

2067 DIAG_PERIOD

180 (s)

AUTO Set the data acquisition period for ILBD within the range of 20

to 65535 (s). Writable only when DIAG_MODE is Stop(0). If the

process fluctuation values are unsteady, change this value to the

longer to enhance the accuracy of the blockage detection.

68

2068 DIAG_PRI

1

AUTO Set alert (DIAG_H_ALM and DIAG_L_ALM) priority for ILBD.

Writable only when DIAG_MODE is Stop(0).

69

2069 DIAG_ERR

0x0000

The following ILBD results are shown.

• A Blocking (bit 2)

• Large Fluctuation of Low Side (bit 3)

• Large Fluctuation of High Side (bit 4)

• Low Side Blocking (bit 5)

• High Side Blocking (bit 6)

• B Blocking (bit 7)

• Invalid Ref BlkF (bit 8)

• Invalid Ref fSPh (bit 9)

• Invalid Ref fSPl (bit 10)

• Invalid Ref fDP (bit 11)

• Outside Diagnosis Range (bit 12)

• Flg Temp Low Alarm (bit 13)

• Flg Temp High Alarm (bit 14)

Bit 15 is used to link the alarm to status of PRIMARY_VALUE,

SECONDARY_VALUE, and TERTIARY_VALUE.

70

2070 DIAG_H_ALM

Alerts for high-pressure side are indicated.

• A Blocking

• B Blocking

• Large Fluctuation of High Side

• High Side Blocking

• Invalid Reference value (BlkF, fSPh, or fDP)

• Outside Diagnosis Range

71

2071 DIAG_L_ALM

Alerts for low-pressure side blockage are indicated due to the

following factors.

• A Blocking

• B Blocking

• Large Fluctuation of Low Side

• Low Side Blocking

• Invalid Reference value (BlkF, fSPl, or fDP)

• Outside Diagnosis Range

72

2072 DIAG_OPTION

0x08FC

AUTO The alarm status linked to an analog alert and LCD display is

selected by DIAG_OPTION in the SENSOR Transducer block.

The alarm selected in DIAG_OPTION is corresponding to bit of

DIAG_ERR.

BIT 15 is used to link the alarm to the status of PRIMARY_VALUE,

SECONDARY_VALUE, and TERTIARY_VALUE.

If set to 1, the signal status becomes “UNCERTAIN: Non Specific”

when the ILBD algorithm detects the abnormality.

Writable only when DIAG_MODE is Stop(0).