<6. Explanation of Basic Items>

6-3

IM 01C25T02-01E

Procedure for automatic adjustment

The procedure for automatic adjustment is as

follows:

F0602.ai

(1) Change the Block mode

Set MODE_BLK to O/S.

(2) Enter zero% pressure

Apply the actual zero% pressure to EJX.

(6) Change the operation mode

Set MODE_BLK to AUTO for normal operation.

(3) Perform zero adjustment

When the value of the applied pressure is written to
CAL_POINT_LO, EJX calculates the amount of
adjustment for zero adjustment. The calculated amount
of adjustment is incorporated into CAL_DEVIATION_LO.

(5) Performance of span adjustment

When the value of the applied pressure is written to
CAL_POINT_HI, EJX calculates the amount of
adjustment for span adjustment. The calculated amount
of adjustment is incorporated into CAL_DEVIATION_HI.

(4) Input pressure for span adjustment

For span adjustment, apply the actual 100% pressure to
EJX.

F0603.ai

0

0

Calculated Value

Calculated Value

Span adjustment

point

After zero

adjustment

After zero adjustment

After zero/span

adjustment

Input pressure

Input pressure

Before zero

adjustment

CAUTION

Span adjustment is a function for adjusting

the gradient with respect to the point of zero

adjustment. If that point is not zero, the gradient

of input and output values cannot be accurately

adjusted. Perform span adjustment after zero

adjustment is completed.

Diagnosis of adjustment results:
When the amount of zero adjustment or span

adjustment exceeds any of the following adjustment

limits, “AL50: Adjustment range error for pressure/

differential pressure” is displayed.
The conditions for zero/span adjustment are as

follows:

<1> The point of zero adjustment is within

±10% of the capsule range.

<2> The amount of zero adjustment is within

±10% of the capsule range.

<3> The point of span adjustment is within the

capsule range.

<4> The input and output gradients after span

adjustment are within ±10% of the gradient at

factory setting.

6.3.4 Functions Relating to Static

Pressure

The static pressure measurement function is unique

to differential pressure transmitters. The pressure

transmitter does not have this function.

Reference to static pressure value:
In SECONDARY_VALUE and TERTIARY_VALUE,

it is possible to refer to the H-side static pressure

value and the L-side static pressure value,

respectively. The update period of these values is

100 msec. The pressure unit is selected by XD_

SCALE.unit in the AI block, in which SECONDARY_

VALUE and TERTIARY_VALUE are selected. The

status is normally Good-Non Specific. However, in

the case of a sensor failure or out of measurement

range, it turns to Bad or Uncertain. For specifics,

refer to Table 8.9.

Setting of signal damping for static pressure:
SP_VALUE_FTIME is parameter that set signal

damping for static pressure signals. The unit of the

signal damping is second.