Yokogawa EJX930A User Manual

Page 45

Advertising
background image

<6. Explanation of Basic Items>

6-4

IM 01C25T02-01E

Determination of the range limit of static

pressure signal:

When the static pressure signal exceeds the

maximum working pressure (MWP) of the capsule,

set the status of PRIMARY_VALUE to Uncertain-

Subnormal. Also, set the status of SECONDARY_

VALUE and TERTIARY_VALUE to Uncertain-

Sensor Conversion not Accurate. The status under

normal conditions is Good-Non Specific.

Adjustment of static pressure signal:
The zero/span adjustment function can be used

for static signals, as in the case of pressure

/differential pressure signals. However, the zero-

point adjustment by the external screw function is

not supported for static pressure signal.
Automatic adjustment

The value of the static pressure exerted on the

point where adjustment is to be made is assigned

to parameters (CAL_SP_POINT_LO, CAL_SP_

POINT_HI). After this assignment is set, the amount

of adjustment is calculated by the transmitter and

automatically updated. The procedure for adjusting

the static pressure signal is the same as that of

pressure/differential pressure signal.
Manual adjustment

From the exerted pressure and the output of the

transmitter, calculate the amount of zero/span

adjustment manually and assign the calculated

value to parameters (CAL_SP_DEVIATION_LO,

CAL_SP_DEVIATION_HI).

Selection of static pressure signal type:
Differential pressure transmitters measure the

absolute static pressure. Based on the assumption

that atmospheric pressure is constant, these

transmitters can output signal equivalent to gauge

pressure calculated by deducting the atmospheric

pressure (fixed value) from the measured absolute

pressure. The type of static pressure signal output

to SECONDARY_VALUE and TERTIARY_VALUE

is selected in SP_VALUE_TYPE. SP_VALUE_

TYPE allows selection of gauge pressure or

absolute pressure. When selecting gauge pressure,

set the value of the atmospheric pressure (fixed

value) to ATM_PRESS. The default value of

ATM_PRESS is equal to the value of standard

atmospheric pressure (101.325 kPa).

6.3.5 Functions Relating to Capsule and

Amplifier Temperature

Reference to capsule and amplifier temperature

value:

In CAP_TEMP_VAL and AMP_TEMP_VAL, it

is possible to refer to the capsule and amplifier

temperature value and status. The update period

of this value is about 1 second. The temperature

unit is selected by XD_SCALE.unit of the AI block,

in which CAP_TEMP_VAL is selected. The status

is normally Good-Non Specific. However, in the

case of sensor failure or out of measurement range,

it turns to Bad or Uncertain. For specifics, refer to

Table 8.9.

Determination of the range limit of capsule

temperature:

The range of capsule temperature is from –40 to

120°C. When the measured capsule temperature

deviates from –50 to 130°C range, set the status of

CAP_TEMP_VAL to Uncertain-Sensor Conversion

not Accurate. The status under normal conditions

is Good-Non Specific. Also, set the status of

PRIMARY_VALUE, SECONDARY_VALUE and

TERTIARY_VALUE TO Uncertain-Subnormal.

Determination of the range limit of amplifier

temperature:

The range of amplifier temperature is from –40 to

120°C. When the measured amplifier temperature

deviates from –50 to 95°C range, set the status of

AMP_TEMP_VAL to Uncertain-Sensor Conversion

not Accurate. The status under normal conditions is

Good-Non Specific.

Advertising