2 status of each parameter in failure mode, Status of each parameter in failure mode -5 – Yokogawa EJX930A User Manual

Page 62

Advertising
background image

<8. Device Information>

8-5

IM 01C25T02-01E

8.2 Status of Each Parameter in Failure Mode

Following tables summarize the value of EJX parameters when LCD display indicates an Alaram.

Table 8.9

Action of each parameters in failure mode related Resource block and Sensor Transducer block

ALARM

Display Cause of Alarm

Resource

block

SENSOR Transducer block*

BLOCK_ERR BLOCK_ERR XD_ERROR PV.STATUS

SV.STATUS

TV.STATUS

CAP_TEMP_

VAL.STATUS

AMP_TEMP_

VAL.STATUS

AL.01

CAP.ERR

Pressure

Sensor problem

other

I/O Failure

BAD:Sensor

Failure

BAD:Sensor

Failure

Capsule

Temperature

Sensor problem

other

I/O Failure

BAD:Sensor

Failure

Capsule

memory

problem

Lost NV Date

BAD:Device

Failure

BAD:Device

Failure

BAD:Device

Failure

BAD:Device

Failure

AL.02

AMP.ERR

Amplifier

Temperature

Sensor problem

other

Electronics

Failure

BAD:Sensor

Failure

Amplifier

memory

problem

Lost NV Date

Lost Static

Data

BAD:Device

Failure

BAD:Device

Failure

BAD:Device

Failure

BAD:Device

Failure

Amplifier

problem

other

Electronics

Failure

BAD:Device

Failure

BAD:Device

Failure

BAD:Device

Failure

BAD:Device

Failure

AL.10

PRESS

Input Pressure

is outside

measurement

range limit of

capsule

UNCERTAIN:

Sensor

Conversion

not Accurate

UNCERTAIN

Subnormal

AL.11

ST.PRSS

Static Pressure

exceeds limit

UNCERTAIN

Subnormal

UNCERTAIN:

Sensor

Conversion

not Accurate

AL.12

CAP.TMP

Capsule

Temperature is

out of range

UNCERTAIN

Subnormal

UNCERTAIN

Subnormal

UNCERTAIN:

Sensor

Conversion

not Accurate

AL.13

AMP.TMP

Amplifier

Temperature is

out of range

UNCERTAIN:

Sensor

Conversion

not Accurate

AL.20

NOT.RDY

No

communication

is found with

LAS.

AL.40

RS O/S

Resource Block

is in O/S mode

Out of

Service

BAD:
Non-specific

BAD:
Non-specific

BAD:
Non-specific

BAD:

Non-specific

AL.41

TB O/S

Sensor TB is in

O/S mode

Out of

Service

Out of

Service

BAD:Out of

Service

BAD:Out of

Service

BAD:Out of

Service

BAD:Out of

Service

AL.50

P.SDEV

P.ZDEV

Trimming

range error

for differential

pressure span

other

Calibration

error

UNCERTAIN:

Engineering

Unit Range

Violation

AL.51

SP.SDEV

SP.ZDEV

Trimming range

error for static

pressure span

other

Calibration

error

UNCERTAIN:

Engineering

Unit Range

Violation

*: For FLG_TEMP_VAL.STATUS, refer to A8.3.5.

Table 8.10

Action of each parameters in failure mode related LCD display

ALARM

Display

Cause of Alarm

LCD TB

BLOCK_ERR

XD_ERROR

AL.79

LCD.RNG

LCD.CFG

Displayed value exceeds limit or LCD configuration error

other

Configuration Error