<4. WIRING>

4-5

IM 01E25D11-01EN

(3) Wiring the Remote Flowtube with AXFA11

Converters

Connect wiring as shown in the figure below.

I+

4-20mA out

Shortbar

I–

CURRENT OUT

AL+

AL–

C

SA

A

B

SB

ALARM OUT

N/–

L/+

POWER SUPPLY

EX2

EX1

EXCIT ATION

P–

SI1+

SI2+

COM

P+

PULSE OUT

STATUS IN

SIGNAL

SO1+

COM

SO2+

STATUS OUT

FUSE

2.5A 250V

AXFA11 converter

Converter

Remote

flowtube

F0409.ai

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate

the shields corresponding to

SA and SB on the remote

flowtube side.

Remote flowtube

EX2

Excitation

cable

EX1

A

B

C

AXFC dedicated

signal cable

Figure 4.1.9 Wiring Diagram

(4) Grounding

CAUTION

Be sure to connect the protective grounding of

the remote flowtube with a cable of 2 mm

2

or

larger cross section in order to avoid electrical

shock to the operators and maintenance

engineers and to prevent the influence of

external noise.

Connect the grounding wire to the mark.

The grounding should satisfy Class D

requirements (ground resistance, 100 Ω or less).

IMPORTANT

Improper grounding can have an adverse affect

on the flow measurement. Ensure that the

instrument is properly grounded.

The electromotive force of the magnetic flowmeter

is minute and it is easy to be affected by noise. And

also that reference electric potential is the same

as the measuring fluid potential. Therefore, the

reference electric potential (terminal potential) of

the flowtube and the converter also need to be the

same as the measuring fluid. Moreover, that the

potential must be the same with ground.

Be sure to ground according to Figure 4.1.10.

Refer to Figure 3.3.5 or 3.3.6 for wiring to the

flanges of remote flowtube when optional grounding

rings are not used.

• Class D requirements (ground resistance.

100 Ω or less).

600 V vinyl insulated

electric cable

(2 mm

2

or larger)

F0410.ai

With optional grounding

rings.

Without optional grounding rings.

(Available only for metal piping)

Figure 4.1.10 Protective Grounding Terminal Location

(5) Installing the Cover
Install the cover to the flowtube by turning it in the

direction of the arrow as shown below. Tighten

the cover locking screw counterclockwise using a

hexagonal wrench (nominal size 3 mm) to lock the

cover.

F0411.ai

Figure 4.1.11 Installing the Terminal Box Cover