<5. MAINTENANCE>

5-3

IM 01E25D11-01EN

5.4 Troubleshooting

Although magnetic flowmeters rarely require maintenance, failures may occur when the instrument is not

operated correctly. This section describes troubleshooting procedures where the cause of the breakdown is

identified through receiver indication.

5.4.1 No Indication

Is an error being displayed? Check the converter display.

NO

YES

NO

NO

YES

NO

YES

NO

Is an error being

displayed?

Is anything else

being displayed?

YES

YES

Check whether the flowtube

is filled with fluid.

Is the flowtube

full of fluid?

Remove the tube and

inspect the inside of the pipe.

Is foreign

matter adhering to

the walls?

Does the

arrow on the tube match the

flow direction?

Refer to the “Alarm Functions” in

the AXFA11.

Perform either of the following:

a. Reinstall the flowtube in

order to make the arrow and

flow direction match.

b. Reverse the setting for the

flow direction.

Change the piping or the

mounting condition so that

the tube fills with the fluid.

Remove the foreign matter.

START

Contact a service center.

Check the resistance between

the flowtube’s A, B, and C

terminals using a circuit tester.

Is the

resistance between A-C or B-C

100 MΩ or more?

NO

YES

F0503.ai