Optional specifications, Optional specifications -5, Noptional specifications – Yokogawa ADMAG AXW User Manual

Page 36

Advertising
background image

<6. OUTLINE>

6-5

IM 01E25D11-01EN

n

OPTIONAL SPECIFICATIONS

: Available – : Not available

Item

Specifications

Applicable Model

Code

General

Submersible

AXW***G AXW***W

Grounding Rings (*1) JIS SUS304 (AISI 304 SS Equivalent), a set of two rings

GR1

Grounding Rings (*2) SUSF304 (AISI 304 SS Equivalent), a set of two rings with gaskets “B”.

GR2

For District Heating

and Cooling or

Condensation-proof

Urethane resin potting is applied in the terminal box of a remote flowtube. Select JIS

G1/2 for the electrical connections. 30-meter dedicated signal and excitation cables are

pre-wired and waterproof glands with union joints are attached at factory.

DHC

User-specified Signal

and Excitation Cable

Length

Available for the submersible type and a model with optional code DHC. The cable

length is limited up to 200 meters combined with an AXFA11 converter. Following “L,”

specify the cable length in three digits as a multiple of 1 meter (e. g., 001, 002, or 005)

for a length up to 5 meters, or as a multiple of 5 meters (i.e., 005, 010, 015, or the like)

for a length of 5 meters or more. If this optional code is not selected, a 30-meter length

cable is attached.

L***

Mass Unit Setting

The flow rate span, transmission pulse weight, and totalizer display pulse weight can be

set in terms of mass unit.

Specify the density of the process fluid when ordering in addition to the mass flow rate

span, transmission pulse weight (for mass unit), and totalizer display pulse weight (for

mass unit).

When ordering a remote flowtube, parameters for 'Mass Unit Setting' will be set in the

corresponding converter before shipment.

1. Density

a. Available density Numerics:

Specify the numeric within the value of 500 to 2000 kg/m

3

, 4.2 to 16.7 lb/gal, or 31.2

to 124.8 lb/cf. And it can be up to five digits, to a maximum of 32000 ignoring the

decimal point. A fraction is limited to the fourth decimal place.

b. Available density units: kg/m

3

, lb/gal, lb/cf

Example: A water density is about 1000 kg/m

3

. In this case specify “1000kg/m

3

”.

However a density is changed by temperature. Specify the actual density.

(The 1000 kg/m

3

is equivalent to 8.345lb/gal and 62.43lb/cf.)

2. The mass flow rate span, transmission pulse weight, and totalizer display pulse

weight

a. Available density Numerics:

Specify the numeric within the value of 0.0001 to 32000. And it can be up to five

digits, to a maximum of 32000 ignoring the decimal point. A fraction is limited to the

fourth decimal place.

b. Mass Units

Available mass units: t, kg, g, klb, lb

Available time units: /d, /h, /min, /s

Note1: In case of specifying the mass flow span, calculate the volumetric flow span

by the setting density, and specify the available value in the mass flow span.

Note2: In case of transmission pulse weight and totalizer display pulse weight,

specify the mass unit which was specified as the flow unit.

MU

G3/4 Female

Waterproof Glands

Waterproof glands for G3/4 conduits or flexible tubes are attached to the electrical con-

nections. Available only for JIS G1/2 electric connections.

EW

Waterproof Glands

Waterproof glands are attached to the electrical connections. Available only for JIS G1/2

electric connections.

EG

Waterproof Glands

with Union Joints

Waterproof glands with union joints are attached to the electrical connections. Available

only for JIS G1/2 electric connections.

EU

Stainless Steel Tag

Plate

A pendant tag plate of stainless steel (AISI 304 SS equivalent) is provided in addition to

the standard nameplate with the tag number inscribed on it.

Dimension (Height x Width): Approx. 12.5 (4.92) x 40 (15.7) mm (inch)

SCT

Direction Change of

Electrical

Connection (*3)

+90 degrees rotated terminal box to change the direction of the electrical connection.

RA

+180 degrees rotated terminal box to change the direction of the electrical connection.

RB

–90 degrees rotated terminal box to change the direction of the electrical connection.

RC

Material Certificate

Material certificates are provided for pipe, electrodes, flanges, and also optional ground-

ing rings when specified.

M01

*1: This is not available for models with process connection JIS F12 (JIS 75M) of sizes 1100 mm (44 in.) to 1800 mm (72 in.). Refer to the section

on “Mounting Procedures” on necessary gaskets.

*2: This is available only for models with process connection JIS F12 (JIS 75M) of sizes 1100 mm (44 in.) to 1800 mm (72 in.). Refer to the section

on “Mounting Procedures” on necessary gaskets.

*3: Refer to the image below.

Standard

+90 degree rotation

+180 degree rotation

–90 degree rotation

Optional Code RA

Optional Code RB

Optional Code RC

Electrical

Connection

Electrical

Connection

Advertising