6. FUNCTION AND DATA SETTINGS

6.3.2 Display of Internal Totalization Values

• The flow converter can display totalization values by setting the pulse weight.

(1) Setting Totalization Pulse Weight

Example: Display 10 liter output per pulse in a flow rate span of

ٗٗٗ m

3

/h

Display

Description

Select the volume unit for pulse weight using parameter number "E1".

Code

Volume Unit

00

01

02

03

04

05

06

Volume unit used in 10

-9

×

flow rate span

Volume unit used in 10

-6

×

flow rate span

Volume unit used in 10

-3

×

flow rate span

Volume unit used in 1

×

flow rate span

Volume unit used in 10

3

×

flow rate span

Volume unit used in 10

6

×

flow rate span

Number of pulses output per second at 100% of output

Example : When pulses are to be output per liter in a span of

m

3

/h, select "02" since L(liter) = 10

-3

×

m

3

Set the low cut range percentage in parameter "E3".
Range of setting : 0 to 100% (of span)

Select pulse output calculation of instantaneous flow rate or flow rate after
damping. (Use parameter "02" to set damping constant.)

Set pulse weight "10 (L)" in parameter number "E2".

Since parameter number "E2" is a 5-digit data item, scrolling is
necessary to display all the data. Mind the decimal point when
setting are made. (The decimal point can be moved if required.)

Defalut: 3%

Defalut: 01
(Damped flow rate value)

* The “n” item can be opened by entering “55” in parameter number “L2”.