9.3 Trouble Shooting

Although magnetic flowmeters rarely require maintenance, failures occur when the

instrument is not operating correctly. Since a failure is located by troubleshooting the
receiving instrument information. This information will be described below.

9.3.1 No Indication

Check the code indication

(see Page 7-5) and examine

depending on the contents.

Contact your
agency or
the nearest
sales office.

Check that the meter tube is

filled with the liquid.

Change piping or

mouting so that the

meter tube is full of

liquid.

Inspect inside of tube by

removing.

Execute either of the following.

a, Install meter so arrow agree

with flow direction.

b,

Change setting in 14 or B14

"FLOW DIR"

Examine the converter display or LED flashing wheather an error is displayed

NO

NO

NO

NO

NO

NO

NO

YES

YES

YES

YES

YES

YES

Replace the
fuse.

YES

Is an error displayed?

START

Is other display appearing?

Is the fuse blown?

Does the display

reappear by replacing

the fuse?

Does the arrow

on meter tube agree with the

flow direction?

Is the meter tube

full of the liquid?

Is foreign matter

sticking to the wall?

Remove foreign matter.

Contact your agency or the

nearest sales office.