1 zero adjustment using data setting keys, Zero adjustment using data setting keys -2 – Yokogawa ADMAG CA User Manual

Page 67

Advertising
background image

IM 1E8B0-01E

8-2

8. ACTUAL OPERATION

8.1.1 Zero Adjustment Using Data Setting Keys

The following two procedures can be used to perform zero adjustment with the data
setting keys.

(1) Hold Down the INC Key for Two Seconds in Flow Rate Data Display Mode.

Display

Description

Code

Content

Call up the setting mode and set "01" in "C1" (zero adjustment enable / disable).

Hold down the INC key for two seconds in flow rate data display mode.

The display panel shows "-ZEro-".
After about 60 seconds, the zero correction is displayed. Then the meter returns to
data display mode.

00

01

Zero adjustment disabled

Zero adjustment enabled

SET SHIFT INC

Default: 01

%

2 seconds

or longer

(2) Display “C2” and Press the SET Key Twice.

Display

Description

Code

Content

Call up the setting mode and set "01" in "C1" (zero adjustment enable/disable).

Call up "C2" and press the SET key twice (as for data entry).

The display panel shows "-ZEro-".
After about 60 seconds, the zero correction is displayed. Then the meter returns to
data display mode.

00

01

Zero adjustment disabled

Zero adjustment enabled

SET SHIFT INC

Twice

60 seconds
later

(Displays the
zero
correction)