2 error description and countermeasures, Error description and countermeasures -5 – Yokogawa ADMAG CA User Manual

Page 70

Advertising
background image

IM 1E8B0-01E

8-5

8. ACTUAL OPERATION

8.2.2 Error Description and Countermeasures

02

03

04

05

08

09

16

17

18

19

06

µ

P FAULT

EEPROM FAULT

A/D(H) FAULT

A/D(L) FAULT

VEL. SAPN>10m/s

VEL. SAPN<0.3m/s

P.SPAN<0.0001p/s

T.SPAN>1000p/s

T.SPAN<0.0001p/s

4-20 LMT ERROR

Microprocessor error

EEPROM error

A/D converter (high frequency side) error

A/D converter (low frequency side) error

Setting for span flow velocity exceeds 11 m/s

Setting for span flow velocity is 0.2 m/s or below

Pulse output rate exceeds 1100 p/s, at 50% DUTY
Pulse output rate exceeds 1000 p/s, at 0.5 ms
pulse width

Pulse output rate exceeds 500 p/s at 1 ms
pulse width

Pulse output rate exceeds 25 p/s at 20 ms
pulse width

Pulse output rate exceeds 15 p/s at 33 ms
pulse width

Pulse output rate exceeds 10 p/s at 50 ms
pulse width

Pulse output rate exceeds 5 p/s at 100 ms
pulse width

Pulse output rate is 0.00005 p/s or below

Internal totalization exceeds 1100 p/s

Internal totalization at 0.00005 p/s or less

4-20 low limit > 4-20 high limit

SIGNAL OVERFLOW

CA

Display on BT

Error Contents

Countermeasures

Check for the following :
● Mixup of the signal, power supply,

execution, and other cable.

● Stray electorical currents in the

measured liquid.

● Improper grouping

Contact the nearest Yokogawa
office, or service center

Contact the nearest Yokogawa
office, or service center

Change setting

Excessive input signal

07

COIL OPEN

Detector coil open-circuit

P.SPAN>1000p/s

10

11

P.SPAN>500p/s

12

P.SPAN>25p/s

13

P.SPAN>15p/s

14

P.SPAN>10p/s

15

P.SPAN>5p/s

Advertising