2 unstable zero, Unstable zero -4 – Yokogawa ADMAG CA User Manual

Page 74

Advertising
background image

IM 1E8B0-01E

9-4

9. MAINTENANCE

9.3.2 Unstable Zero

NO

Examine whether or not the
meter tube is filled with liquid
and the liquid does not contain
bubbles.

Examine whether or not the

valves leak.

Recheck liquid property.

Replace valves.

Sufficiency shield the meter.
Locate the meter as far as
possible from this equipment.

NO

NO

NO

YES

NO

NO

NO

YES

YES

YES

YES

YES

YES

It is convenient to provide a gas
vent hole on the downstream
side of the meter to check for
bubbles.

In particular, be careful for
horizontal meter mouting,.
Change mounting position or
change to vertical mounting
with upward flow so that the
tube is completely filled with
liquid.

Mount a deforming device on

the upstream side of the meter

· Confirm the meter grounding.
· Confirm whether or not the

grounding wire is
disconnected from the earth
flange.

a, Insure that the liquid

conductivity does not change
beyond fluid limits.

b, Do not use the meter with

liquid conductivity below the
specified value.

c, Decrease flow velocity near

the conductivity low limit as
much as passble.

• Remove dirt and scale on

and near electrodes.

• Change to vertical meter

mounting.

• Change the measuring mode
to the standard mode.

START

Are bubbles contained in the

liquid?

Is the meter tube filled with

liquid completely?

Are valves leaking?

• Does the fluid

conductivity vary or is
it easy to change?

• Is Liquid conductivity

sometimes below the
specified value?

Is the liquid likely to cause

deposit?

Is the flowmeter located

near mortors, transformers, or

other electric equipment?

*

*

Is grounding complete?

Insulation deterioration is suspected.
Examine the meter.

Grounding resistance must be
100

or less.