6.3.3 Presetting for Totalization Display

• This function is used to reset or preset totalization values of the display unit.
• Hold down the SHIFT key for more than 2 seconds while the totalization values of

the flow rate reading are displayed to set the totalization value to the value set in
parameter number "E5".

Display

Description

Select totalization enable in parameter number "E4"

Code

Description

00

01

Totalization presetting inhibit

Totalization presetting enabled

Set the totalization preset value in parameter number "E5". The initial value
is 0, if it is no setting, the function is zero setting.

Hold down the SHIFT key for more than 2 seconds while the totalization
values of the flow rate reading are displayed to set the totalization value to
the value set in parameter number "E5"

Default: 00 (disabled)

Default: 0

SET SHIFT INC

6.3.4 Damping Time Constant

• The time constant can be changed by setting the parameter No. “02” to suppress a

fluctuation or change a response time.

• The time constant influences to flow rate, pulse output and internal totalization.

However, in case “00” is selected in parameter No. “n1”, the pulse output and internal

totalization are not influenced by it.

Display

Description

Set the value in parameter number "02".
Range of possible settings : 1.0 to 200.0 seconds.

Default: 3 seconds