2 setting parameters, 1 sensor configuration, Setting parameters -4 – Yokogawa Wireless Temperature Transmitter YTA510 User Manual

Page 11: Sensor configuration -4

Advertising
background image

IM 01C50T03-01E

3-4

3. OPERATION

3.2 Setting Parameters

3.2.1

Sensor Configuration

When the sensor type or the number of wire connec-
tions changes, the following parameters must be reset.

Sensor type setting;

D10: SENSOR1 TYPE, E10: SENSOR2 TYPE

Wire connections setting;

D20: SENSOR1 WIRE, E20: SENSOR2 WIRE

Figure 3.1 diagram shows the wire connections to the
input terminals of the transmitter and sensor type
selections for the parameters in each connection case.
Note that TCs and mV are categorized as Group A and
RTDs and ohm as Group B.

Check the connections between the input terminals and
temperature sensors and set the correct sensor type and
the number of wire connections for the parameters.

Thermocouple and DC voltage (TC & mV)

Resistance thermometer(RTD) and resistance (2-wire type)

1-input model YTA110, YTA310, and YTA320

2-input model YTA320

1
2
3
4
5

1
2
3
4
5

1
2
3
4
5

1
2
3
4
5

(–)

(+)

(B)

(A)

(B)

(B)

(A)

(B)

(A)

(A)

(B)

Thermocouple(TC) &
Resistance thermometer(RTD) and resistance (3-wire type)

1
2
3
4
5

(–)

(+)

(+)

F0301.EPS

Sensor1

Sensor1

Sensor1

Sensor1

1
2
3
4
5

(B1)

(A1)

(B2)

(A2)

Sensor2

Sensor1

1
2
3
4
5

(B1)

(B1)

(A1)

(B2)

(B2)

(A2)

Sensor2

Sensor1

Sensor2

Sensor1

1
2
3
4
5

(–)

(+)

(A)

(B)

(B)

Sensor2

Sensor1

Resistance thermometer(RTD) (4-wire type)

Resistance thermometer(RTD) and resistance (3-wire type)

Thermocouple and DC voltage (TC & mV)

Resistance thermometer(RTD) and resistance (2-wire type)

Resistance thermometer(RTD) and resistance (3-wire type)

Group A

Group B

Group B

*

Group B

Group A

Group A

Group A

Group B

Group B

Group B

Group B

Group B

TYPE B (IEC584)
TYPE W3 (ASTM988)
TYPE W5 (ASTM988)
TYPE E (IEC584)
TYPE J (IEC584
TYPE K (IEC584)
TYPE L (DIN43710)
TYPE N (IEC584)
TYPE R (IEC584)
TYPE S (IEC584)
TYPE T (IEC584)
TYPE U (DIN43710)

Pt100 (IEC751)
Pt200 (IEC751)
Pt500 (IEC751)
JPt100 (JIS)
Ni120 (STI INC)
Cu (SAMA RC21-4)
ohm

mV

TYPE B (IEC584)
TYPE W3 (ASTM988)
TYPE W5 (ASTM988)
TYPE E (IEC584)
TYPE J (IEC584
TYPE K (IEC584)
TYPE L (DIN43710)
TYPE N (IEC584)
TYPE R (IEC584)
TYPE S (IEC584)
TYPE T (IEC584)
TYPE U (DIN43710)

Pt100 (IEC751)
Pt200 (IEC751)
Pt500 (IEC751)
JPt100 (JIS)
Ni120 (STI INC)
Cu (SAMA RC21-4)
ohm

[

* Only for 2 or 3-wire type]

mV

Group A

Group B

*

: Without ohm

Figure 3.1

Input terminal wire connection diagram and sensor type categories

Sensor type selection

Thermocouple

TYPE W3, W5 (ASTM988)
TYPE B, E, J, K, N, R, S, T
(IEC 584)
TYPE L, U (DIN 43710)

Resistance thermometer

Pt100, Pt200, Pt500
(IEC 751)

[2-, 3- or 4-wire]

JPt100 (JIS)
Ni120 (STI INC), Cu
(SAMA RC21-4)

DC voltage

mV

Resistance

ohm [2- or 3-wire]

Advertising
This manual is related to the following products: