IM 01C50T03-01E

3-12

3. OPERATION

Pin position of SW2

Write Protect Status

Y

N

No

F0355.EPS

W P

Yes

Y

N

W P

SW2

SW2

SW2
Hardware Write Protect

SW1
Burnout output direction
upon hardware failure

CPU Assembly

Figure 3.3

SW2 setting

3.2.14 Sensor Trim

Each YTA transmitter is factory-characterized based on
the standard sensor curve and uses the information to
produce a process variable output. The sensor trim
function is used to adjust to the transmitters internal
interpretation of the input signal.

Input

ZERO

One point trim

Input

ZERO

GAIN

Two points trim

F0310.EPS

Output

Output

Figure 3.4

Trim function images

(a) Zero/Gain Point Adjustment

J07: IN TRIM MODE

J10: SNSR1 ZERO, J20: SNSR1 GAIN

K10: SNSR2 ZERO, K20: SNSR2 GAIN ...YTA320 only

1) Connect the temperature transmitter and the

calibration device as shown in Figure 3.5 and warm
up for at least three minutes.

a. Wiring of power supply and output

b. Example of wiring of thermocouple
or DC voltage input (1-input type)

F0311.EPS

+ Output signal

Load resistance

DC voltage generator

or thermocouple

Voltmeter

1
2
3
4
5

c. Example of wiring of thermometer resistor
4-wire type (1-input type)

Variable resistor

or thermometer resistor

1
2
3
4
5

(+)

(A)

(B)

(B)

(A)

(–)

Figure 3.5

Example of wiring for calibration equipment

2) Check the sensor type with the D: parameter.
3) Select the input trimming mode in J07:IN TRIM

MODE. The following selections are offered.

V.R./ZERO&GAIN
V.R./ZERO
TEMP/ZERO&GAIN
TEMP/ZERO

Select “V.R./ZERO&GAIN” or “V.R./ZERO” when
the calibration device is DC voltage generator or
Variable resistor, or select “TEMP/ZERO&GAIN”
or “TEMP/ZERO” when the device is Temperature
sensor.

4) Perform zero-point adjustment.

• When the input trimming mode is “V.R./

ZERO&GAIN” or “V.R./ZERO”, apply the value
for the zero-point shown in the Table 3.1 depend-
ing on the specified sensor type. Wait until the
input from the calibration device becomes stable.

• When the input trimming mode is “TEMP/

ZERO&GAIN” or “TEMP/ZERO”, expose the
temperature sensor to calibration temperature for
the zero-point. Wait until the input from the
temperature sensor becomes stable.