Wiring Procedure (Panel Mount Type)

The power supply terminals and protective ground terminals are located on the rear
panel.

1.

Turn OFF the power switch on the recorder and open the power terminal cover.

2.

Wire the power cord and the protective ground cord to the power supply terminals.

Use ring-tongue crimp-on lugs (designed for 4-mm screws).

3.

Close the power supply terminal cover and secure it with the screw.

The proper torque for tightening the screws is 0.6 N•m.

Screw for fixing

the power terminal

cover in place

Power

terminal block

Power

terminal cover

Open

Philips

screwdriver

Crimp-on lug

with insulation

sleeves

Power cord

Protective ground

cord

• Other than /P1 • /P1

(24 VDC/AC power supply)

L

N

L

N

DC/AC24V

+

-

Wiring Procedure (Portable Type (/H5x Option))

1.

Check that the power switch to the instrument is turned OFF.

2.

Connect the plug on the accessory power cord to the power supply connector on

the rear panel.

100-240V~

50/60Hz

40V

A MAX

3.

Ensure that the power outlet to be used meets the conditions on the previous

page and that the voltage of the power supply is within the maximum voltage
rating of the power cord, then connect the other end of the power cord to the
power supply outlet.
The AC outlet must be of a three-prong type with a protective earth ground termi-
nal.

Turning ON/OFF the Power Switch

The power switch is located inside the door at the lower right. The power switch is a push
button.
Press once to turn it ON and press again to turn it OFF. When the power switch is turned
ON, a self-diagnosis program runs for a few seconds, and the recorder is ready for
operation.