Setup example (3) of 0 to 10 v input – Yokogawa µR10000 User Manual

Page 36

Advertising
background image

36

IM 04P01B01-02E

Selecting the Unit

16.

Use the

CHARACTER

key and

key to set

unit character and then press the

key. (For the procedure, see “Entering Characters” on page 23. For the characters
that can be used, see “Explanation” on page 38).

Unit:

CHR:%-

Finishing the Unit Setting

17.

When

Setting complete screen is displayed, press the ESC key.

03-03 Channel

Setting complete

18.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Setup Example (3) of 0 to 10 V Input

Set channel 04 to 0 to 10 V input and measure the range 0.0 to 400.0 m

3

/h. The 20 V DC

voltage range is used. The unit is converted using the linear scaling function. The scaling
range is –20000 to 30000 excluding the decimal point.
The measured values in the range of 0.0 to 400.0 m

3

/h are recorded in a width of 100

mm on the chart paper.

–20 V

20 V

Linear scaling Measurable range of 20 V

0.0 m

3

/h

400.0 m

3

/h

(Scale left)

100 mm

(Scale right)

0.00 V

10.00 V

(Span left)

(Span right)

Chart paper

Entering Setting Mode

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

Selecting the Range

2.

Press the key with

Range shown on the screen.

Set=Range

Input range and

Displays a description of the setup item.

Selecting the Channel Range

3.

Press the

key to set the first channel to

04, and then press the key.

First channel Last channel

CH=04-04

First channel 01-

Displays the selectable range of channels.

4.

Likewise, set the last channel to

04 and then press the key.

* When the key, key, or

CHARACTER key is pressed while holding down the SHIFT key, the

operation is reversed as when the respective key is pressed by itself.

Setting the Input Range and Alarm on Measurement Channels