Yokogawa µR10000 User Manual

Page 35

Advertising
background image

35

IM 04P01B01-02E

Selecting the Input Type

5.

Press the

key to select

1-5V and then press the key (see “Explanation”

on page 38).

Mode=1-5V

Scales and record

Setting Span Left

6.

Set Span_L to

1.000 and press the key.

Press the

key to select the desired digit. Press the

key to select the

value.

Span_L= 1.000

0.800/ 1.200V

Displays the range of Span_L.

Setting Span Right

7.

Likewise, set Span_R to

5.000 and press the key.

Span_R= 5.000

4.800/ 5.200V

Displays the range of Span_R.

Setting the Decimal Position and Scaling Left

8.

Display Scale_L.

Scale_L= 0.00

-200.00/ 300.00

Displays the scaling range.

9.

Press the

key to select the desired digit.

Scale_L= 0.00

-200.00/ 300.00

10.

Press the

key to select space and then press the key (Scale_L is set to

0.0).

Scale_L= 0.0

-200.00/ 300.00

Setting Scaling Right

11.

Likewise, set Scale_R to

500.0 and press the key.

Scale_R= 500.0

-2000.0/ 3000.0

Displays the scaling range.

The

Setting complete screen is displayed. When this screen is displayed, the

settings entered up to then are applied.

Finishing the Settings

12.

When

Setting complete screen is displayed, press the ESC key.

The

Set=Range screen is displayed.

03-03 Channel

Setting complete

Set=Range

Input range and

Setting the Unit

13.

Press the

key to select

Unit and then press the key.

Set=Unit

Engineering unit

Displays the description of setting item.

14.

Press the

key to set the first channel to

03 and then press the key.

First channel Last channel

CH=03-03

First channel 01-

Displays the selectable range of channels.

15.

Likewise, set the last channel to

03 and then press the key.

* When the key or key is pressed while holding down the

SHIFT key, the operation is reversed as

when the respective key is pressed by itself.

Setting the Input Range and Alarm on Measurement Channels