In the Module the IR Detector/Filter Temperature is monitored to set

the zero request bit as follows:
Zero Request if a ±5°C temperature change has occurred.
Ambient Temperature Out Of Range is set when the IR

Detector/Filter Temperature is not within its allowed range.

Sample Cell Temperature:

The Module sample cell temperature is monitored by a thermistor in

contact with the sample cell. This temperature measurement is used

as follows:
An “out of range” sample cell temperature condition causes Ethanol,

CO

2

and CO data to be declared “data invalid.”

Sample Cell Temperature Out Of Range is set when the Sample Cell

Temperature is not within its allowed range.

NO

Sensor Temperature:

A thermistor located in each NO sensor assembly monitors NO

sensor temperature. It is used as follows:
Sensor response during NO gas concentration measurement.

An “out of range” NO sensor temperature condition causes NO data

to be declared “data invalid. ”The NDIR Pressure Measurements.

1.4.5 The NDIR Pressure Measurements.

The Module includes two pressure transducers for measurement of

the pressures. An absolute pressure transducer measures the

pressure in the sample cell. A differential pressure transducer on the

power section of the Printed Circuit Board (PCB) measures the

difference between ambient pressure and the sample gas pressure

between the inlet port and the pump. These pressure measurements

are used as follows:
Sample Cell Pressure (absolute). Sample cell pressure is used as

follows:
Gas law compensation” during Ethanol, CO and CO

2

gas

concentration measurement.
Zero/O

2

Span and Span calibration (via the use of temperature

compensated Ethanol, CO

2

and CO data).