Figure 4.10. The Main Menu Screen displaying Zero Auto mode.

4.3 Span Calibration

4.3.1 Introduction.


The Multi Gas Monitor – INNOVA 1316A-3 is designed to be span-stable.

Frequent span calibration is not required to maintain accurate gas concentration

measurements. The Multi Gas Monitor may need to be routinely span calibrated

with suitable mixtures of calibration gases. Span calibration is recommended at

the following intervals:
¨

The VM0102A module measuring Ethanol, CO, and CO

2

: Once every 12

months.

¨

NO sensor VM0201A: After replacement of a NO sensor, and once every

month during normal operation.

¨

O

2

Sensor VM0200A: Span calibration is performed during every Zero/O

2

Span Calibration execution.

¨

When any span calibration has failed.


4.3.2 Checking and Changing the Particle Filters

It is recommended to change the inlet filter DS2306 before

performing any calibration.