3 calibration gases – LumaSense Technologies INNOVA 1316A-3 User Manual

Page 66

Advertising
background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6024-12

1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 66 of 96

Figure 4.11.The particle filter DS2306

The particle filter is attached to the rear panel of the monitor.

Dismantle the filter from the luerlok fittings and replace it with a new

DS2306.

4.3.3 Calibration Gases

To perform a full calibration gases are required.

Special gas mixtures containing known concentrations of Ethanol,

Carbon Monoxide and Carbon Dioxide are required to span calibrate

the VM0102A

If UA1372A is mounted with electrochemical sensors O

2

(VM200A)

and NO (VM0201A) then calibration gases to span calibrate the

sensors are also needed.

The gas mixture shown in

Table 4.1

is allowed concentration ranges

for calibrating the monitor

Gas

Concentration Range

Ethanol

100 ppm to 1 %

Oxygen

1% to 25 %

Carbon Dioxide

1 % to 20 %

Carbon Monoxide

0.5% to 15 %

Nitrous Oxide

100 ppm to 5000 ppm

Nitrogen

Balance

Table 4.1 Allowed gas concentration ranges for 1316A-3 Span calibration