1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 70 of 96

Figure 4.16 Main window displaying measurement started..

Open the status window and press the “Zero”-button in the Main window, in

order to perform a Zero calibration.

The ambient air is taken from the Zero reference inlet on the rear panel during

Zero calibration. In the Status window observe that the “Proc in prog”-status is

“In Progress”, while Zero calibration. is done, see

figure 4.17

.

When Zero calibration has successfully finished the “Proc in prog”-status is “Not

in Progress”, see

figure 4.18

.