LumaSense Technologies INNOVA 1316A-3 User Manual

Page 78

Advertising
background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6024-12

1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 78 of 96

After SPAN calibration revert to default mode 5. In the “Mode Select”-window set

the group “Action” to “Temporary Change” and the group “New Mode” to “5 –

6800, Single-SOL, New”, See figure 4.28. Press the “Send” button to set mode

5.

Figure 4.28 Select Mode 5 in the Mode Select window

Stop the measurement by pressing the “Stop”-button in the Main window. Finally

set the group “Zero” to "Auto" in the Main window, see figure 4.29

Figure 4.29 Set Zero calibration to Auto

This concludes the Span calibration.

4.3.6 To perform a span calibration with single component gas.

If only one gas is to be span calibrated this can be performed using a

single component gas in Nitrogen balance.

Before doing the Span calibration please make sure that Zero

calibration is set to “Manual” and the mode is set to 3 as described in

section 4.3.5

.