1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 56 of 96

Figure 4.2. The Zero O

2

span calibration window to select Purge time during a

calibration.

Using the Window above the BZ6014 can extend the Purge Time (PT) in

1-second increments.

2.

Power on.

The first Zero during a power on cycle takes 5 seconds longer than a

subsequent Zero with the same PT value.

3.

NO Concentration

When the Zero/O

2

Span Calibration command is transmitted, and the NO

concentration is above a set value. The purge time is extended in order to

completely purge the NO sensor.

4.2.1 Zero Calibration Procedure


The CO

2

, CO, Ethanol, and NO channels are zero calibrated and the O

2

channel is span calibrated. Using the selected Purge Time (PT) (see

figure 4.2 ).

Assuming that the monitor and the PC are connected and the

software is running (displaying the Main Menu), the zero calibration

continues as follows: