Correct weighing – Seca 985 User Manual

Page 27

Advertising
background image

Model 985

27

GB

Checking that the contents of the calibration counter are

correct

This seca scale is calibrated. Calibration may be performed only by authorised centres.

To ensure this, the scale is fitted with a calibration counter which records any change to

data relevant to calibration.The calibration counter stamp can be found on the display

housing.
If you want to check whether the scale is properly calibrated, proceed as follows:
– If necessary, switch off the scale.
– Keep any key depressed and start the

scale.
The current contents of the calibration

counter is flashing on the display for a

few seconds.

– Compare the output contents of the

calibration counter with the number

shown on the calibration counter

sticker.
For valid calibration, both numbers must

match. If sticker and calibration counter

do not match, then re-calibration is re-
quired. Contact your service agent or

seca Customer Service.

Correct weighing

– Press the Start key.

se(A

,

8.8.8.8.8.

and

0.0

appear consec-

utively in the display. The scale is then

automatically set to zero and ready to

use.
The plug symbol

appears in the dis-

play.
The scale remains in permanent opera-
tion until it is switched off.

– Get the patient into bed.
– Read patient weight off the digital dis-

play.
If a load greater than 250 kg is placed

on the scale,

St0P

appears in the dis-

play.

– To switch off the scale, press the Start

key again.

Both numbers must match.