… the display

Er:1:11

appears?

– At least one weighing cell is not correctly connected or the connection is back to front.

Check the connections. The colours of plug and socket must match.

Switch the scale off using the Start key and restart. The scale then works normally

again. If this is not the case, inform Servicing.

… the display

Er:1:12

appears?

– The maximum weight for switching on was exceeded when the scale was switched

on, e.g. if bed and patient are being weighed. Remove the load from the scale and start

it again. The scale then works normally again. If this is not the case, inform Servicing.

… one segment is permanently illuminated or is not illuminated at all?
– The relevant segment is faulty. Inform Servicing.

8. Maintenance

Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating

plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 27).
We recommend having your scale serviced by your local service agent before having it

recalibrated. seca Customer Service will be pleased to give you further assistance.

9. Technical data

Dimensions
width:

540 mm

height:

990 mm

depth:

540 mm

weight

approx. 25 kg

Measuring feet
width:

328 mm

height:

150 mm

depth:

212 mm

weight:

approx. 6.2 kg

Cable length:

2.80 m

Size of figures in display

20 mm

Temperature range

+10 °C to +40 °C

Calibration

as per Directive

2009/23/EC

approved for medical use,

Class III

Medical product as per Directive

93/42/EEC

Class I with measuring function

Maximum load

250 kg