Moving and storing the scale

– Switch off the scale.
– Unlock the actuating lever of the meas-

uring foot by pressing down the round

knob and folding out the lever.

– Raise the actuating lever to lower the

castor.

– Pull the measuring foot out to the side,

wind up the cable and place the meas-
uring foot in one of the compartments

of the equipment trolley.

– Proceed in exactly the same way to re-

move the remaining three measuring

feet and store them on the trolley.

– Disconnect the mains plug from the

socket and likewise store the mains unit

on the equipment trolley.

– Release the brakes on the equipment

trolley.

You can now move the bed scale to its

next point of use or “park” it somewhere

out of the way until it is required again.

5. Setting damping

Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display re-
acts. You can set damping to suit requirements.
– Press the

FUNC key and keep it de-

pressed for 2 seconds.

FIL and the current setting appear in

the display:
0 = slight damping

s fast determination of weight

1 = average damping

s normal determination of weight

2 = strong damping

s slow determination of weight.

– Press an arrow key if you want to

change the set value.

– Press the

FUNC key again to store the

displayed value. The scale is then back

in normal weighing mode.

The settings are retained even once the scale is switched off.