6.3 Function tables

6.3.1 Menu A, quick setup

INFORMATION!
Depending on the device version, not all functions are available.
The description "PF option" used in the tables only refers to the TIDALFLUX 4000 measuring
sensor / "CAP option" only refers to the OPTIFLUX 7000 measuring sensor.

No.

Function

Settings / descriptions

A1 language

A1

language

Language selection depends on the device version.

A2 Tag

A2

Tag

Measuring point identifier (Tag no.) appears in the LC display header.

A3 reset

A3

reset

-

A3.1

reset errors

reset? Select: no/yes

A3.2

reset counter 1

reset counter? Select: no / yes (available if activated in C5.9.1)

A3.3

reset counter 2

reset counter? Select: no / yes (available if activated in C5.9.2)

A3.4

reset counter 3

reset counter? Select: no / yes (available if activated in C5.9.3)

A4 analogue outputs (only for HART

®

)

A4

analogue outputs

Applicable to all current outputs (terminals A, B and C), frequency outputs

(terminals A, B and D), limit switches (terminals A, B, C, and/or D) and the

1st display page / line 1.

A4.1

measurement

1) Select: volume flow / mass flow (not

not

not

not valid for PF (partly filled)) / diagnosis

value / flow speed / coil temperature / conductivity (not

not

not

not valid for PF (partly

filled)) and CAP (capacitive)) / level (only

only

only

only valid for PF (partly filled))

2) Use for all outputs? (also use this setting for Fct. A4.2...A4.5!)

Setting: no (applies only to the main current output ) / yes (applies to all

analogue outputs)

A4.2

unit

Selection of the unit from a list, depending on the measurement.

A4.3

range

1) Setting for main current output (range: 0...100%)

Setting: 0…x.xx (format and unit, depending on measurement, see A4.1 and

A4.2 above)

2) Use for all outputs? Make setting, see Fct. A4.1 above !

A4.4

low flow cutoff

1) Setting for main current output (sets output value to "0")

Setting: x.xxx ± x.xxx% (range: 0.0…20%)

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

2) Use for all outputs? Make setting, see Fct. A4.1 above !

A4.5

time constant

1) Setting for main current output (applicable to all flow measurements)

Setting: xxx.x s (range: 000.1…100 s)

2) Use for all outputs? Make setting, see Fct. A4.1 above !