Electrical connections – KROHNE IFC 300 Converter EN User Manual
Page 45

ELECTRICAL CONNECTIONS
4
45
IFC 300
www.krohne.com
08/2010 - 4000069803 - MA IFC 300 R04 en
4.5.2 Connecting the signal and field current cables, wall-mounted housing
• The outer shield of signal cable A and/or B is connected via the stranded drain wire.
• If a shielded field current cable is used, the shield must NOT
NOT
NOT
NOT be connected in the housing of
the signal converter.
• Bending radius: ≥ 50 mm / 2"
1 Open the housing cover.
2 Pass the prepared signal cable through the cable entry and connect the corresponding strand-
ed drain wires and conductors.
3 Connect the stranded drain wire of the outer shield.
4 Pass the prepared field current cable through the cable entry and connect the corresponding
conductor.
Any shield that is present must NOT
NOT
NOT
NOT be connected.
5 Tighten the screw connections of the cable entry and close the housing cover.
Figure 4-14: Electrical connection of the signal and field current cables, wall-mounted housing
INFORMATION!
Ensure that the housing gasket is properly fitted, clean and undamaged.