TECHNICAL DATA
160
IFC 300
www.krohne.com
08/2010 - 4000069803 - MA IFC 300 R04 en
8.5 Measuring accuracy (except TIDALFLUX)
Reference conditions
• Medium: water
• Temperature: 20°C / 68°F
• Pressure: 1 bar / 14.5 psi
• Inlet section: ≥ 5 DN
X [m/s]: flow velocity
Y [%]: deviation from the actual measured value (mv)
1
2
3
DN [mm]
DN [inch]
Accuracy
Curve
OPTIFLUX 5300
10….100
3/8…4
0.15% of mv + 1 mm/s
1
150...300
6...12
0.2% of mv + 1 mm/s
2
OPTIFLUX 2300 / 4300 / 6300
10….1600
3/8…80
0.2% of mv + 1 mm/s
2
OPTIFLUX 1300
10…150
3/8…6
0.3% of mv + 2 mm/s
3
OPTIFLUX 2300 / 4300
>1600
>64
0.3% of mv + 2 mm/s
3
OPTIFLUX 4300 / 5300 / 6300
<10
<3/8
0.3% of mv + 2 mm/s
3
OPTIFLUX 7300
25...100
1...4
v ≥ 1 m/s / 3.3 ft/s:
±
0.5% of mv
-
v < 1 m/s / 3.3 ft/s:
±
0.5% of mv + 5 mm/s
WATERFLUX 3300
25...600
1...24
0.2% of mv + 1 mm/s
2