Operation – KROHNE IFC 300 Converter EN User Manual

Page 130

Advertising
background image

6

OPERATION

130

IFC 300

www.krohne.com

08/2010 - 4000069803 - MA IFC 300 R04 en

C5.2 display

C5.2

display

-

C5.2.1

language

Language selection depends on the device version.

C5.2.2

contrast

Adjust display contrast for extreme temperatures.

Setting: -9…0…+9

This change takes place immediately, not just when setting mode is exited!

C5.2.3

default display

Specification of the default display page that is returned to after a short

delay period.

Select: none (the current page is always active) / 1. meas. page (show this

page) / 2. meas. page (show this page) / status page (show only status

messages) / graphic page (trend display of the 1st measurement)

C5.2.4

self test

Not available at this time.

C5.2.5

SW.REV.UIS

Serial no. of the circuit board, version no. of user software, production date

of the circuit board

C5.3 and C5.4 1. meas. page and 2. meas. page

C5.3

1. meas. page

…

stands for 3 = meas. page 1 and 4 = meas. page 2

C5.4

2. meas. page

C5.….1

function

Specify number of measured value lines (font size)

Select: one line / two lines / three lines

C5.….2

measurement 1.line

Specify measurement for 1st line.

Select: volume flow / mass flow (not

not

not

not valid for PF (partly filled)) / diagnosis

value / flow speed / coil temperature / conductivity (not

not

not

not valid for PF (partly

filled)) and CAP (capacitive)) / level (only

only

only

only valid for PF (partly filled))

C5.….3

range

0…100% of the measurement set in Fct. C5.….2

0…xx.xx _ _ _ (format and unit depend on the measurement)

C5.….4

limitation

Limitation before applying the time constant.

xxx%; range: -120…+120%

C5.….5

low flow cutoff

Sets output value to "0"

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C5.….6

time constant

Range: 000.1…100 s

C5.….7

format 1.line

Specify decimal places.

Select: automatic (adaptation carried out automatically) /

X (= none) …X.XXXXXXXX (max. 8 digits)

C5.….8

measurement 2.line

Specify measurement 2.line (only available if this 2.line is activated)

Select: bar graph (for the measurement selected in the first line) / volume

flow / mass flow (not

not

not

not valid for PF (partly filled)) / diagnosis value / flow

speed / counter 1 / counter 2 / counter 3 / conductivity (not

not

not

not valid for PF

(partly filled) and CAP (capacitive)) / coil temperature / operating hours /

level (only

only

only

only valid for PF (partly filled))

C5.….9

format 2.line

Specify decimal places.

Select: automatic (adaptation carried out automatically) /

X (= none) …X.XXXXXXXX (max. 8 digits)

C5.….10

measurement 3.line

Specify measurement 3.line (only available if this 3.line is activated)

Select: volume flow / mass flow (not

not

not

not valid for PF (partly filled)) / diagnosis

value / flow speed / coil temperature / conductivity (not

not

not

not valid for PF (partly

filled) and CAP (capacitive)) / counter 1 / counter 2 / counter 3 / operating

hours / level (only

only

only

only valid for PF (partly filled)) / current input A / current input

B

Advertising