Electrical connections – KROHNE IFC 300 Converter EN User Manual

Page 38

Advertising
background image

4

ELECTRICAL CONNECTIONS

38

IFC 300

www.krohne.com

08/2010 - 4000069803 - MA IFC 300 R04 en

4.4.7 Preparing field current cable C, connection to signal converter

• Field current cable C is not part of the scope of delivery.
• Bending radius: ≥ 50 mm / 2"

Required materials:

• Shielded 3-wire copper cable with suitable heat-shrinkable tubing
• DIN 46 228 wire end ferrules: size according to the cable being used

Length and cross-section of field current cable C

DANGER!
A non-shielded three-wire copper cable is sufficient for the field current cable. If you
nevertheless use shielded cables, the shield must NOT

NOT

NOT

NOT be connected in the housing of the signal

converter.

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

Length

Cross-section A

F

(Cu)

[m]

[ft]

[mm

2

]

[AWG]

0...150

0...492

3 x 0.75 Cu

1

3 x 18

150...300

492...984

3 x 1.50 Cu

1

3 x 14

300...600

984...1968

3 x 2.50 Cu

1

3 x 12

1 Cu = copper cross-section