Measurement: Cyanuric Acid

C401/102 Colorimeter

16

Add content of one
Cyanuric Acid Reagent Sachet

Shake for
1 minute

9.

Gently push the vial back fully into the sample well of the instrument,
whilst again aligning the ▼ mark on the vial with the ▲ mark on the
meter.

10. Wait for two minutes and then press the READ/ENTER key within

five minutes from the time the Cyanuric Acid reagent was added to
the vial.

11. The meter will flash [CYNA] whilst the measurement is taken. It will

then display the Cyanuric Acid concentration in parts per million
(ppm).

Cyanuric
Acid
Sample

READ

ENTER

ppm