Calibration: Chlorine, Free & Total, Chlorine Dioxide, Bromine, Ozone

C401/301/201/103/104/105 Colorimeter

42

16. Press the READ/ENTER key. The meter now accept the calibration

value and momentarily display [HoLD] and CAL annunciator before
returning to the beginning of the calibration as in step 3.

17. To continue with another calibration, press the MODE key to select

the desired calibration mode (for multi-parameter colorimeter).

18. To exit calibration mode, press ZERO key to return to the

measurement mode.

CAL

MODE

READ

ENTER

Press Mode to select
another calibration test

ZERO

Press ZERO to return
to measurement mode

Chlorine Free & Total calibration sequence

NOTE: If you wish to abort the calibration
process, do NOT press the READ/ENTER key as
in step 16. Press the ZERO key instead. The
meter will display [AbRT] and CAL annunciator to
signify that the calibration has been aborted
without saving the calibration value. The meter
will then return to the beginning of the user
calibration as in step 3.

NOTE: If an error occurs during calibration, the display will present an error
message. Please refer to Section 4

Troubleshooting Guide for details.