Thermo Fisher Scientific C101 Colorimeter User Manual

Page 30

Advertising
background image

Measurement: Chlorine Dioxide

C103 Colorimeter

26

Acid to the sample before carrying out the above test procedure. Use a
multiplication factor to correct for the dilution of the sample.

Chlorine:

Chlorine levels greater than 6ppm may interfere. Use two sachets of
Chlorine Dioxide (Glycine) Reagent in step 5) to remove the interference.

Hardness:

No effect at less than 2000ppm CaCO

3

.

Metals and Chlorine in Combination:

If chlorine is present, some metals may interfere with the test. Use two
sachets of Chlorine Dioxide (Glycine) Reagent in step 5) to remove the
interference.

Monochloramine:

Monochloramine will cause a gradual increase in the Chlorine Dioxide
result over time. Take the test result within 1 minute of adding the reagent
sachet to avoid these errors.

Oxidants, including Bromine and Iodine:

These will interfere with the test method at all levels. It is recommended
that an alternative test method is used if these species are present.

Ozone:

May cause interference at levels above 1.5ppm.

Oxidised Manganese or Oxidised Chromium:

All levels will cause interference. To remove this interference:

1.

Test a 10ml sample as described in the above procedure.

2.

Adjust the pH of a separate 10ml sample to pH6-7 using 1N Sodium
Hydroxide or 1N Sulphuric Acid.

3.

Add 1 drop of 20% Potassium Iodide Solution; swirl to mix and then
wait 60 seconds.

4.

Add 3 drop of 5g/l Sodium Arsenite Solution and swirl to mix.

5.

Use this as the sample in the above test procedure and make a note
of the result. Use a multiplication factor to correct for the dilution of
the sample.

Advertising