Measurement: Chlorine Dioxide

C103 Colorimeter

24

Chlorine
Dioxide
Sample

READ

ENTER

ppm

Add content of one

Chlorine Dioxide (Glycine)

Reagent Sachet

Swirl vial approximately

20 seconds until powder

dissolves

7. Unscrew the cap from the vial and add into it the content of a

Chlorine, Free (DPD) Reagent Sachet.

8. Replace the cap, ensuring it is screwed on finger tight to the vial.

Swirl the vial repeatedly for approximately 20 seconds, until the
powder has dissolved.

Note: A small amount of the powder may remain un-dissolved in the vial, but
this will not affect the measurement. Do not shake or invert the vial as this
may result in the loss of Chlorine Dioxide.

Swirl vial approximately

20 seconds until powder
dissolves

Add content of one

Chlorine, Free (DPD)

Reagent Sachet

9. Gently push the vial back fully into the sample well of the meter,

whilst again aligning the ▼ mark on the vial with the ▲ mark on the
meter.

10. Press the READ/ENTER key within one minute from the time the

Chlorine, Free (DPD) Reagent Sachet was added to the vial.

11. The meter will flash [CLO2] whilst the measurement is taken. It will

then display the Chlorine Dioxide concentration in parts per million
(ppm).