Calibration: Cyanuric Acid

C401/102 Colorimeter

44

Shake for

1 minute

Cyanuric
Acid
Sample
70 ppm

ppm

READ

ENTER

CAL

CAL

MODE

READ

ENTER

Press Mode to

select another

calibration test

ZERO

Press ZERO to

return to
measurement

mode

CAL

11. Select another clean vial, un-screw the cap and fill it with 10 ml of the

prepared Cyanuric Acid standard solution (fill up to the line on the
vial).

12. Select a Cyanuric Acid Reagent Sachet. Open the sachet and tip the

contents into the vial containing the calibration standard.

13. Replace the cap, ensuring it is screwed on finger tight

to the vial. Shake the vial for approximately one
minute, until the powder has dissolved.

Note: a small amount of the powder may remain un-dissolved in the vial, but
this will not affect the measurement.

14. Insert the vial fully into the sample well of the meter, whilst again

aligning the ▼ mark on the vial with the ▲ mark on the meter.

15. Wait for two minutes, then press the READ/ENTER key within five

minutes from the time the Cyanuric Acid reagent was added to the
vial.

16. The meter will display [--Rd--] whilst the calibration standard is being

measured and then displays [SAVE] to prompt for calibration
acceptance and saving of calibration data.

17. Press the READ/ENTER key. The meter will now accepts the

calibration value and momentarily display [HoLD] and CAL
annunciator before returning to the beginning of the calibration as in
step 3.

18. To continue with another calibration, press the MODE key to select

the desired calibration mode (for multi-parameter colorimeter).

19. To exit the calibration mode, press the ZERO key to return to the

measurement mode.