Restoring Factory Calibration
C401/301/201/101/102/103/104/105 Colorimeter
50
The meter is now ready for measurement
READ
ENTER
READ
ENTER
Press MODE and
keys simultaneously and
hold for 2 seconds before
releasing
ON/OFF
Meter’s Model Number
Meter’s Revision Number
Measurement Selection Mode
Press MODE
key to select
or
YES
No
To restore factory
calibration
To retain last
calibration values
without retrieving
factory calibration.
METER OFF
OFF
ON
MODE
MODE
Figure 9: Restoring Factory Calibration