Thermo Fisher Scientific C101 Colorimeter User Manual
Page 59
Specifications
C401/301/201/101/102/103/104/105 Colorimeter
55
Measurement Method
Photometric
Light Source
Light Emitting Diode (LED)
Wavelength 525
nm
Detector Silicon
photodiode
Absorbance Range
0 – 2.5 Abs
Photometric Precision
+/- 0.0015 Abs
Calibration Points
User selectable; 1 point per colorimetric test
Display
4-digit 14-segments customized liquid
crystal display with annunciators
Sample Cells
Borosilicate glass with screw caps, fill line
and indexing mark
Height x diameter: 51 x 25 mm ( 2 x 1 in)
Sample Required
10 ml (0.33 oz)
Operating Temperature
Range
0 – 50
°C (32°F to 122°F)
Sample Temperature
Range
0 – 50
°C (32°F to 122°F)
Operating Humidity Range
0-90% RH non-condensing at 30
°C (86°F)
Power Supply
4 AAA alkaline batteries
Battery Life
> 3000 tests
Electromagnetic
Compliance (EMC)
Emitted Interference - EN 61326
Immunity to Interference - EN 61326
Enclosure Rating
IP67
Insulation Rating
Pollution degree 2
Weight
Instrument: 200 g (7 oz)
Instrument in carrying case: 1.25 kg (2.75
lb)
Dimensions
Instrument: 6.8(W) x 15.5(L) x 4.6(H) cm;
(2.7 x 6.1 x 1.8 in)
Carrying case: 16(W) x 35(L) x 12(H) cm;
(6.3 x 13.8 x 4.7 in)